IEC 61223-3-5 Ed. 2.0 b:2019

Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests - Imaging performance of computed tomography X-ray equipment

International Electrotechnical Commission , 09/16/2019

Publisher: IEC

File Format: PDF

$208.00$417.00


IEC 61223-3-5 - defines the essential parameters which describe the performance of the CT scanners with regard to image quality, patient dose and positioning; - defines the methods of testing the essential parameters; - evaluates compliance with the tolerances of the parameters specified by the accompanying documents. These methods rely mainly on non-invasive measurements, using appropriate test equipment, performed during the installation or after it has been completed. Signed statements covering steps in the installation procedure may be used as part of the acceptance test report. This part of IEC 61223 is intended to assist in performing the acceptance tests on a CT scanner. The aim is to verify compliance of the installation with specifications affecting the image quality, patient dose and positioning.

IEC 61223-3-5 Ed. 2.0 b:2019 History

IEC 61223-3-5 Ed. 2.0 b:2019
IEC 61223-3-5 Ed. 1.0 b:2004

More IEC Standards PDF

IEC 62616 Ed. 1.0 b:2010

IEC 62616 Ed. 1.0 b:2010

$72.00 $145.00

ICC CA-EC-2010

ICC CA-EC-2010

$24.00 $49.00

IEC 61753-086-6 Ed. 1.0 b:2010
IEC 60728-11 Ed. 3.0 en:2010